AN-109 Verifying Continuity with fiber MAP

Share This...

AN-109 Verifying Continuity with fiber

Overview

OPL-MAX has a built in sequence to test continuity for multi-fiber cables. The only requirement is that the user has a multichannel source and a multichannel power meter.

Startup

Navigate to the OPL-MAX software shortcut located either in the Start menu or on the desktop. When the splash screen appears, select the proper setup via the Selected Setup drop down menu.

Figure 1: OPL-Max splash screen
Figure 1: OPL-Max splash screen
Figure 2: Error message
Figure 2: Error message

If either the multichannel source or the multichannel power meters are not connected, the software will give an error message and the software must be restarted when both units are connected.

Figure 3: Default Sequence tab
Figure 3: Default Sequence tab
Figure 4: Instruments tab in setup
Figure 4: Instruments tab in setup

With the multichannel units properly connected, the software will open with the correct setup and sequence on the Setup > Sequence (default) tab.

The user can verify the correct units are being recognized as well as their ID. This can be viewed in Setup > Instruments.

Navigate back to the Sequence tab and click on fiberMAP on the top right. A popup screen should appear showing a matrix of inputs versus outputs.

Figure 5: Popup showing a matrix of inputs versus outputs
Figure 5: Popup showing a matrix of inputs versus outputs

Setup

Figure 6: FiberMap setup configuration
Figure 6: FiberMap setup configuration Threshold

Before running the fiberMap one needs to configure the setup.

Threshold

The dBm threshold value is the minimum power level which will be considered a “lit” channel.
The relative dB value creates a range of accepted measurements that indicate a fiber is connected.

Inputs

This is the number of inputs channels from the source.  For a 12 fiber MPO cable this value would be 12.

Outputs

This is the number of output channels to be monitored.

Simulation

When setup is complete, click Explore. The software will run through all outputs per input and read the power levels.

Figure 7: Simulation of hardware setup for verifying continuity with FiberMap
Figure 7: Simulation of hardware setup for verifying continuity with FiberMap
Figure 8: FiberMap output screen before analysis
Figure 8: FiberMap output screen before analysis

Note that the power levels where there is no connection (i.e. A1:B12, A3:B1) read dark values. When the simulation is completed, it will highlight values within the threshold. This indicates a connection from one input to the output.
The highest power is added from all power measurements. For example, -20dBm was measured as the highest power. Therefore, the connection A1:B1 would read 0, while A1:B2 would read 57.

Figure 9: FiberMap output screen after analysis
Figure 9: FiberMap output screen after analysis
Figure 10: Shows an MTP-MTP cable with a 1:1 pinout (input 1 routes to output 1, etc.)
Figure 10: Shows an MTP-MTP cable with a 1:1 pinout (input 1 routes to output 1, etc.)